To assess the subjective image quality in digital panoramic radiograph, when the
tube current level (mA) is reduced.
By Dimax 3 digital planmeca x-ray machine two digital panoramic projections can
obtained for a dry skull , the first projections was performed at standard exposure
setting , the tube current level (mA) was reduced in the second exposure by 25% ,
50% , and 75% respectively . 10 observers evaluated both projections for the
anatomical landmarks which are commonly appearing in the panoramic radiograph.
Radiographs taken at the reduced (mA) level of 75% had highly statistically
significant inferior score P<0.0001 for anatomical details.
At reduced tube current level (mA), radiographs had inferior subjective image quality,
but a reduction in tube current level (mA) of approximately 50% is accepted.
Downloads
Published
Issue
Section
License
The Journal of Mustansiria Dental Journal is an open-access journal that all contents are free of charge. Articles of this journal are licensed under the terms of the Creative Commons Attribution International Public License CC-BY 4.0 (https://creativecommons.org/licenses/by/4.0/legalcode) that licensees are unrestrictly allowed to search, download, share, distribute, print, or link to the full texts of the articles, crawl them for indexing and reproduce any medium of the articles provided that they give the author(s) proper credits (citation). The journal allows the author(s) to retain the copyright of their published article.